Compounding & Resolving Prism Calculator

Combine a horizontal and a vertical prism into one resultant prism and base angle.

Resultant prism from horizontal and vertical

Enter Prism Vectors
Resultant Vector Output

Compounded Resultant Prism

Resultant Prism Magnitude (R): 5.00 Δ Base Angle: 53.1°
R = √(3.00² + 4.00²) = 5.00 Δ at 53.1°

The formulas

R = √(H² + V²)

θ = arctan(V ÷ H)

H, V
Horizontal and vertical prism in prism dioptres.
R
The single resultant prism.
θ
Base angle measured from the horizontal, 0° to 90°. The calculator gives this angle; you place it in the right quadrant.

To split a prism back into its parts (resolving), work backwards by hand: H = R × cos θ and V = R × sin θ. For example, 5 Δ at 30° from the horizontal is 4.33 Δ horizontal and 2.50 Δ vertical.

Worked example

Right eye: 4.00 Δ base out and 3.00 Δ base up. With H = 4.00 and V = 3.00 the calculator shows 5.00 Δ and Base Angle: 36.9°. Base up and out on a right eye is the 90–180° quadrant of the 360° chart, so the order reads 5.00 Δ at 180 − 36.9 = 143°. The same prism on a left eye would be at 37°.

Placing the angle on the 360° chart

The 360° notation is read as the practitioner faces the wearer, with 0° to the practitioner's right for both eyes. For the right eye 0° is base in; for the left eye 0° is base out. Up is 90° and down is 270° for both.

Base directionRight eye (OD)Left eye (OS)
Up and inθ180° − θ
Up and out180° − θθ
Down and in360° − θ180° + θ
Down and out180° + θ360° − θ

When it is used

  • Writing a lab order when a prescription has prism in both directions: a lens can carry only one prism, so the two are combined.
  • Checking a finished lens: on a lensmeter the target moves diagonally, by the resultant, not by two separate amounts.
  • Splitting prism between the eyes to even out thickness, after resolving it into its horizontal and vertical parts.

Limits. The calculator takes magnitudes only, from 0 to 15 Δ in 0.25 Δ steps, and does not know the eye or the base directions. It adds prisms as vectors, which is accurate for the small amounts used in spectacles. See the medical disclaimer.

Questions

Why can't the lab grind the two prisms separately?

A lens has one front and one back surface, so it can be tilted only one way. A horizontal and a vertical prism together make one tilt with one base direction: the resultant.